6000-5D VOCs Component Online Analyzer

The 6000-5D is designed for continuous real-time online monitoring of C2~C12 volatile organic compounds in ambient air and fenceline (covering 108 PAMS and EPA TO-15 target compounds) with a single FID or dual FID coupled with Deans Switch.

    Technical Data

    Detector

    FID

    Measurement range

    0.1 ppb to ppm level

    Maximum temperature rise rate

    1000/min

    Adsorption/desorption temperature

    -10~350℃(adjustable

    Sampling rate

    5~120ml/minadjustable

    Accuracy

    ±10%

    Precession

    RSD5%

    Gas

    Ultrapure N2H2 or Air

    Data interface

    R~232, Ethernet Interface

    Operating environment

    110V/60Hz or 220V/50Hz±10% with max. 3.5kW, 10~40/20%~90%RH operating range

     

    Features

    The Nutech innovative and specially improved comprehensive separation column was used to meet the separation and analysis needs of 108 target compounds, including PAMS and EPA TO-15.

    The original low-temperature preconcentration technology greatly improves the detection limit, meets the detection requirements of VOC components in low or ultra-low concentration air, and can report the concentration of TVOC at the same time.

    The application of the new generation of advanced technology enables the system to maximize the separation of moisture in samples and effectively avoid FID quenching.

    With FID detector, it has good response, wide linear range, and automatic real-time self-test function.

    Using MFC to accurately control the flow rate to effectively ensure the accuracy of test results.

    PC control, real-time data acquisition and transmission, automatic recording, storage, and support QC/QA reportWith historical data (including atlas) query function, the system can automatically save and restore data if the power is cut off unexpectedly.